Siglent SMM3311X SMM3312X
Specification conditions
Temperature 23 °C ±5 °C
Humidity 30% to 80% RH
Calibration period 1 year
Measurement speed 1 PLC (power line cycle)
After 60 minutes of warm-up Ambient temperature change is less than ±3 °C after self-calibration execution
Outputs
Quadrants 4
Max. acquisition rate 100,000 points/s
Number of output channels 1 2
Max. output/sink power 31.8 W
DC or pulsed ±210 V ≤0.105 A
±21 V ≤1.515 A
±6 V ≤3.03 A
Pulsed only ±200 V ≤1.515 A
±6 V ≤10.5 A
Maximum current limitation
Max current limitation: For 21 V/1.515 A and 6 V/3.03 A ranges, the total max current is limited by the table below for using 2 channels. Max current is not limited to using only 1 channel.
I1: CH1 current
I2: CH2 current
I1(CH1:±6V) + I2(CH2:±6V) ≤4A
I1(CH1:±6V) + I2(CH2:±6-21V)X1.6 ≤4A
I1(CH1:±6-21V) + I2(CH2:±6V)X0.625 ≤2.5A
I1(CH1:±6-21V) + I2(CH2:±6-21V) ≤2.5A